ARTHROCARE CORPORATION EVAC 70 XTRA HP COBLATOR II; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
|
Back to Search Results |
|
| Catalog Number |
EIC5874-01 |
| Medical Device Problem Code |
Arcing (2583)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/27/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Event or Problem Description
|
|
It was reported that, during an ent procedure, one (1) evac 70 wand was sparking.The procedure was resumed, using an equivalent s+n backup, with a delay greater than 30 min.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).H11: h3, h6: the reported device was received for evaluation.A visual inspection observed the returned instrument shows no manufacturing abnormalities.Product was out of the original packaging.No packaging returned.The tip is heavily worn from use.There are signs of arc/burn damage on the shaft near the tip spacer.A functional evaluation was performed on the returned device and the opened device was tested and plugged into the controller and registered settings (7,3).Activating the wand, almost no energy is produced at the tip and the wand struggles to activate on both coblate and coag.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A definitive root cause for the reported issue could not be determined.Factors that could have contributed to the failure include procedural variance, or deviations from established protocols, increasing risks and leading to unintended consequences.Other factor lines include 1) used non-conductive media.2) ensure that the wand tip (including return electrode) is completely surrounded by conductive media during use.Based on this investigation, the need for corrective action is not indicated.
|
| |
|