|
The cobas e602 analyzer serial number was not provided.The cobas e602 analyzer had been deinstalled and disassembled.Therefore, further investigation was not possible.A general analyzer or reagent problem was not present, because the qc prior to the event was within ranges.The investigation was unable to determine a specific root cause.
|
|
There was an allegation of questionable elecsys hcg+ss results from the cobas e602 analyzer on (b)(6) 2026, a patient sample was tested on the cobas e602 analyzer with a result of 193.1 miu/ml.On (b)(6) 2026, the clinician indicated that the result was inconsistent with the patient's status, and a new sample was drawn.Using a cobas pro e801, the result for the new sample was <0.1 miu/ml.Using a cobas pro e801, the original sample was repeated, and the result was <0.1 miu/ml.
|