INSTITUT STRAUMANN AG BL, Ø 3.3MM NC, SLACT. 12MM, RXD, GUIDED; ENDOSSEOUS DENTAL IMPLANT
|
Back to Search Results |
|
| Catalog Number |
021.2212G |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
Fibrosis (3167)
|
| Date of Event |
10/09/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The batch number could be verified.Our manufacturing q-system assures, that production and process controls are in place to ensure that batches confirm to the applicable specifications before they are distributed.The fracture of a dental implant is a known long-term complication of endosseous dental implants.The manufacturers trend analysis confirms that the reported implant fracture rate associated with its dental implants is low and remains consistent with the experience as documented in the literature.There are different reasons why implants break and this problem has been illuminated by various authors (e.G.Maeglin 1988, beumer 1989, tetsch 1991, worthington 1992).Beumer and lewis 1989 report the following causes: a) production / design flaws b) inadequate fitting of the suprastructure c) occlusal overload d) bruxism e) bone resorption around the implants f) insufficient axle alignment of the implants g) trauma h) biomechanical causes i) design of the suprastructure.
|
| |
|
Event or Problem Description
|
|
The clinician reports the implant was inserted (b)(6) 2017 in ada 5.On (b)(6) 2025, fracture of the implant was verified.Patient presented with fair oral hygiene.The device was forwarded to the manufacturer.At the event the patient experienced: mobility.No further patient complications were reported.
|
| |
|
Search Alerts/Recalls
|
|
|