Catalog Number TSVT4B10
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/26/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the mount of the implant at tooth site #14 would not come off.It was noted that the fixture could not be removed.It was extracted immediately after placement and another fixture was placed.The dentist requested an investigation into why it was do difficult to remove.
 
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).A2: age and date of birth unknown / not provided.A3: patient sex unknown / not provided.A4: weight unknown / not provided.Customer has indicated that the product is in process of being returned to zimvie for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
This report is being submitted to relay additional information and device evaluation.The following sections are being reported: b4: date of this report was updated.D9: device availability and return date were updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 10, 3331, 4109 and 4111.H6: investigation findings code was added: 180.H6: investigation conclusions code was added: 4307.H11: narrative/data was updated.Zimvie received the reported device for evaluation.Visual evaluation / functional test was performed, implant and mount shows signs of usage with markings and dried blood.The mount did not disengage/release from implant.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root cause determined from the investigation was torque applied during placement/seating exceeds recommended value.Therefore, based on the available information, a device malfunction did occur.The implant and mount did not disengage/release from each other.The reported event is confirmed.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMP,TSV,4.1,10,MTX,MG
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key24404750
Report Number0001038806-2026-00912
Device Sequence Number15509702
Product Code DZE
UDI-Device Identifier00889024019829
UDI-Public(01)00889024019829(17)270721(10)1256654
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101977
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 03/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberTSVT4B10
Device Lot Number1256654
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/20/2026
Supplement Report FDA Received Date03/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/22/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
Patient EthnicityNon Hispanic
-
-