| Model Number |
67940068P2 |
| Medical Device Problem Code |
Unintended System Motion (1430)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/16/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is unavailable for evaluation.The root cause cannot be determined at this time.A supplemental report will be submitted upon receipt of additional information to support this investigation.
|
| |
|
Event or Problem Description
|
|
The surgeon used 2 spinal cages for anterior cervical discectomy and fusion (acdf).The surgeon reported that the cage had come out forward 2 days after surgery, he should perform surgery again.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation revealed that there was a report relating to two (2) modulus cervical, 5x19x16mm 7 (67940068p2) spacers.It was reported that a patient underwent an acdf on (b)(6) 2025.Two (2) days post-operatively, it was reported that a spacer had migrated anteriorly.No additional adverse effects were reported, and the device(s) remain in-situ, however a revision surgery was planned/suggested.No radiographs were provided and the complaint could not be confirmed.Labeling, dhr, complaint history, and risk reviews were completed with no deficiencies, discrepancies or adverse trends noted.Complaint monitoring will continue and additional action will be taken in the event that an adverse trend is identified.No additional action is planned at this time.
|
| |
|
Event or Problem Description
|
|
The surgeon used 2 spinal cages for anterior cervical discectomy and fusion (acdf).The surgeon reported that the cage had moved forward 2 days after surgery, he should perform surgery again.This event occurred in japan.No other adverse effect reported to patient.
|
| |
|