• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED LONG APPLIER; CLIP, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. LIGACLIP*MCA MED LONG APPLIER; CLIP, IMPLANTABLE Back to Search Results
Catalog Number MCM30
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/27/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4) date sent: 2/20/2026 d4: batch # a9704m.Investigation summary the product was returned to ees for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcm30 device was received with a clip in jaws and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and formed one (1) malformed clip; then in the next actuations, no clips were fed into the jaws even though the device was being actuated in several occasions.In order to evaluate the condition of the device¿s internal components, the device was disassembled.Upon disassembling, the camplate was found to be broken causing the malformed clip and eighteen (18) clips were found inside clip track.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Device history review a manufacturing record evaluation was performed for the finished device batch a9704m number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that during the unk surgery.After clamped the clips and clip could not close.Changed another one to complete the procedure.There was no patient consequence reported, no additional information could be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
orla o'mahony
475 calle c
guaynabo 
*  
3035526892
MDR Report Key24405477
Report Number3005075853-2026-01513
Device Sequence Number8159056
Product Code GDO
UDI-Device Identifier10705036002482
UDI-Public(01)10705036002482(17)291031(10)488D46
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K820837
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberMCM30
Device Lot Number488D46
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2026
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2026
Initial Report FDA Received Date02/20/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/21/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-