(b)(4) date sent: 2/20/2026 d4: batch # a9704m.Investigation summary the product was returned to ees for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcm30 device was received with a clip in jaws and with no apparent damage.In an attempt to replicate the reported incident, the device was tested for functionality.Upon testing, the device was cycled and formed one (1) malformed clip; then in the next actuations, no clips were fed into the jaws even though the device was being actuated in several occasions.In order to evaluate the condition of the device¿s internal components, the device was disassembled.Upon disassembling, the camplate was found to be broken causing the malformed clip and eighteen (18) clips were found inside clip track.As part of ees quality process all devices are manufactured, inspected, and released to approved specifications.No conclusion could be reached as to what may have caused the reported incident.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Device history review a manufacturing record evaluation was performed for the finished device batch a9704m number, and no non-conformances were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|