| Model Number |
VICM5_12.1 (-6.0) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
11/21/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported that following an implantable collamer lens (icl) implantation procedure, the patient experienced refractive surprise.The lens was later exchanged for a same size, different power lens and the problem was resolved cause of the event was unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
4110: a lens work order search has been performed: no similar complaints within associated lots were found.Manufacturer narrative: secondary surgical intervention to remove and replace the lens are identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: d4.Claim# (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
10: returned product evaluation - the subject lens was received dry within a microcentrifuge tube.Visual inspection found an optic and haptic deformed lens.Dimensional and functional inspections found the lens to be within specifications.Claim# (b)(4).
|
| |
|