| Catalog Number |
471858 |
| Medical Device Problem Codes |
Leak/Splash (1354); Reflux within Device (1522)
|
| Health Effect - Clinical Code |
Hemorrhage/Blood Loss/Bleeding (1888)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility: event 1: "the product is back flowing terribly and making a mess.I ordered the standard bore and small bore, and the same problem happened with both.I've used 3 out of each box to make sure it wasn't just one bad set.A nurse also tried and had the same issue." "it's back flow of blood.I disconnected the line, clamped it, and the blood came flowing out of end of the extension set onto the patients clothing.".
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
| |
|
Search Alerts/Recalls
|