Catalog Number 471858
Medical Device Problem Codes Leak/Splash (1354); Reflux within Device (1522)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Based on the data from the investigation, the root cause of the reported incident was unable to be determined.The reported defect was unable to be confirmed.A review of the device history record (dhr) was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Additional Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event or Problem Description
As reported by the user facility: event 3: "the product is back flowing terribly and making a mess.I ordered the standard bore and small bore and the same problem happened with both.I've used 3 out of each box to make sure it wasn't just one bad set.A nurse also tried and had the same issue.It's back flow of blood.I disconnected the line, clamped it, and the blood came flowing out of end of the extension set onto the patients clothing.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARESAFE¿
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer (Section G)
B. BRAUN MEDICAL INC.
824 12th ave
bethlehem PA 18018
Manufacturer Contact
elena reifenstein
824 12th ave
bethlehem, PA 18018
8334251464
MDR Report Key24406323
Report Number2521402-2026-00219
Device Sequence Number13237972
Product Code FPA
UDI-Device Identifier04046955238279
UDI-Public(01)04046955238279(11)230525(17)260430(10)0061884062
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K191658
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2026
Device Catalogue Number471858
Device Lot Number0061884062
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/02/2026
Supplement Date Received by Manufacturer02/02/2026
Initial Report FDA Received Date02/20/2026
Supplement Report FDA Received Date02/27/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/25/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-