| Catalog Number |
515628 |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/30/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the bd spike set c180-o had leakage.The following information was provided by the initial reporter: it was reported that the drug leaked out from puncture site of spike while administering soldem.
|
| |
|
Additional Manufacturer Narrative
|
|
If a device evaluation and/or device history review is completed, a supplemental report will be filed.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) follow up mdr for device evaluation: one iv bag assembled with a spike connector was provided to our quality team for evaluation.Functional testing was performed and a leakage between the spike and iv bag was observed, noting the stopper of the iv bag is torn.Additional evaluation determined that the spike diameter was too large for use with this iv bag.All products undergo visual and functional inspections throughout the manufacturing process to ensure device quality and performance, including verification that all critical dimensions meet specification.A device history review was conducted for lot 2507506, and no deviations or nonconformances related to the reported concern were identified.All records confirmed the product met specifications, with no issues noted.Bd has reviewed this report and identified a potential trend related to this concern.While we have confirmed that our product is manufactured to approved specifications, a project has been initiated to further investigate and address this matter.Complaints received for these devices and reported condition will continue to be tracked and trended.Our quality team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd spike set c180-o had leakage.The following information was provided by the initial reporter: it was reported that the drug leaked out from puncture site of spike while administering soldem.The rubber stopper seems to have been dislodged from the side of the puncture site, so the leak was from the insertion site.
|
| |
|