• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W. L. GORE & ASSOCIATES, INC. GORE® MOLDING & OCCLUSION BALLOON CATHETER; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W. L. GORE & ASSOCIATES, INC. GORE® MOLDING & OCCLUSION BALLOON CATHETER; CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY Back to Search Results
Catalog Number MOB37
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code Perforation of Vessels (2135)
Date of Event 01/27/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A.2.Patient dob/age was asked but not available.H.6.Investigation findings code c21: conclusions pending results of product history review.W.L.Gore & associates, inc.(gore) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by gore, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Blank fields present on this report include required fields and fields determined to be not applicable.Blank required fields indicate that the information was not provided, was deemed unavailable or was not applicable.This report does not constitute an admission or a conclusion by fda, gore, or its associates that the device, gore or its associates caused or contributed to the event described in the report.In particular, this report does not constitute a legal admission by anyone that the product described in this report has any defects or has malfunctioned, as defined from a legal standpoint.These words are included in the report and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Event or Problem Description
The following was reported to gore: on (b)(6) 2026, a patient was being treated for an abdominal aortic aneurysm using a gore® excluder® aaa endoprosthesis system.A gore® molding & occlusion balloon was used for ballooning.After deployment of a plc141000 device, the physician ballooned the iliac artery.While the physician was ballooning, something in the iliac appeared to pop on the fluoro, and about 5 seconds later the patient's blood pressure dropped.The physician took an angio and saw that the iliac had ruptured with extravasation.The physician relined the iliac artery with two plc121400 devices.The rupture was resolved, and the patient tolerated the procedure, and was discharged from the hospital (b)(6) 2026.Reportedly, the iliac artery was stenotic and calcified prior to the procedure.
 
Additional Manufacturer Narrative
H.6.Type of investigation code b18: the device was discarded at the treating facility and was therefore not available for analysis.H.6.Investigation findings code c19: the device history record was reviewed to ensure that each manufacturing and inspection operation was performed and indicated the product met creagh medical specifications and procedures.H.6.Investigation conclusions code d15: there is insufficient information available for gore to reasonably draw conclusions related to aspects of the event, therefore conclusion code ¿d15: cause not established¿ is being used.Insufficient information may include limited or missing relevant medical records, involvement of multiple implanted devices (including non-gore devices) in the field of treatment, patient non-compliance, and/or a general lack of available detail or specificity related to an adverse event and/or device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® MOLDING & OCCLUSION BALLOON CATHETER
Common Device Name
CATHETER, INTRAVASCULAR OCCLUDING, TEMPORARY
Manufacturer (Section D)
W. L. GORE & ASSOCIATES, INC.
1505 n. fourth street
flagstaff AZ 86004
Manufacturer (Section G)
MEDICAL PHOENIX 1 B/P
32360 n. north valley parkway
phoenix AZ 85085
Manufacturer Contact
bryce wargin
1505 n. fourth street
flagstaff, AZ 86004
9285263030
MDR Report Key24409092
Report Number3007284313-2026-04585
Device Sequence Number8299931
Product Code MJN
UDI-Device Identifier00733132639489
UDI-Public(01)00733132639489(17)280706
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K172567
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Catalogue NumberMOB37
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2026
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/21/2026
Supplement Report FDA Received Date03/10/2026
Date Device Manufactured07/08/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-