Model Number 1888TC/46
Medical Device Problem Codes Over-Sensing (1438); Pacing Problem (1439)
Health Effect - Clinical Code Syncope/Fainting (4411)
Date of Event 02/01/2026
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the patient was admitted in the hospital for syncope.Upon interrogation, the patient's right ventricular (rv) lead and right atrial (ra) lead exhibited noise episodes.The ra lead exhibited over-sensed noise resulting in inappropriate automated mode switch (ams) episodes.Programming changes were made to have the right ventricular (rv) lead sensitivity adjusted to avoid noise over-sensing.The patient will be continued to be monitored and was in a stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL ST
Common Device Name
No Match
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
navpreet singh
15900 valley view court
sylmar, CA 91342
MDR Report Key24410908
Report Number2017865-2026-03672
Device Sequence Number5939108
Product Code NVN
UDI-Device Identifier05414734501743
UDI-Public(01)05414734501743(10)4177626(17)160731
Combination Product (Y/N)Y
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P960013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2016
Device Model Number1888TC/46
Device Lot Number4177626
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/01/2026
Initial Report FDA Received Date02/22/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured07/19/2013
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PM2272 ASSURITY PACEMAKER.
Patient SexMale
-
-