RESPIRONICS, INC. REMSTAR AUTO A-FLEX; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DE562S
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
 
Event or Problem Description
A remstar auto a-flex device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and observed damaged foam that cannot be disassembled from the chassis.In addition, dust/dirt was found inside the device.Additionally, the patient¿s complaint was not confirmed, and the device was scrapped.
 
Additional Manufacturer Narrative
The manufacturer previously received information of a remstar auto a-flex device returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and observed damaged foam.Additionally, the device was scrapped.Please note that following further review of device evaluation performed by the third-party service center it was determined that no foam degradation was identified within the device.Analysis of past events has not indicated an adverse event has occurred due to the reported allegation or would be likely to recur if the product allegation was to recur.In addition, a review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.This event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury and/or reportable product malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REMSTAR AUTO A-FLEX
Common Device Name
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key24411260
Report Number2518422-2026-103411
Device Sequence Number13413080
Product Code BZD
UDI-Device Identifier00606959030503
UDI-Public(01)00606959030503
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K131982
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberDE562S
Device Catalogue NumberDE562S
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/23/2024
Supplement Date Received by Manufacturer04/28/2026
Initial Report FDA Received Date02/23/2026
Supplement Report FDA Received Date04/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured01/21/2013
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1974-2021
Patient Sequence Number1
Patient SexUnknown
-
-