| Model Number |
DE562S |
| Medical Device Problem Code |
Degraded (1153)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The capa process identified complaints not being created for quality issues found during third party servicing.Based on a review of records impacted by the identified issue, it was determined that this record needed to be corrected and is now being submitted accordingly.
|
| |
|
Event or Problem Description
|
|
A remstar auto a-flex device was returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and observed damaged foam that cannot be disassembled from the chassis.In addition, dust/dirt was found inside the device.Additionally, the patient¿s complaint was not confirmed, and the device was scrapped.
|
| |
|
Additional Manufacturer Narrative
|
|
The manufacturer previously received information of a remstar auto a-flex device returned to a third-party service center in reference to the voluntary field safety notice / recall notification related to the sound abatement foam in certain cpap, bipap, and mechanical ventilator devices.During the evaluation of the device at the third-party service center, the device was visually inspected and observed damaged foam.Additionally, the device was scrapped.Please note that following further review of device evaluation performed by the third-party service center it was determined that no foam degradation was identified within the device.Analysis of past events has not indicated an adverse event has occurred due to the reported allegation or would be likely to recur if the product allegation was to recur.In addition, a review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.This event is not reportable under fda 21 cfr part 803 as it does not meet the criteria for a death, serious injury and/or reportable product malfunction.
|
| |
|
Search Alerts/Recalls
|