| Catalog Number |
68HP-650 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Bacterial Infection (1735)
|
| Date of Event |
01/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "bacterial infection" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: bacterial infection.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿incision and drainage to obtain culture, pseudomonas aeruginosa¿.This record is for the left side.Device was explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Fi summary with the information collected during the investigation, there is enough evidence to support that units involved in this work order were manufactured under controlled conditions in accordance with abbvie¿s procedures and were no defects/problems/issues found in the documents related to the reported event.Seal strength was successfully completed and results were acceptable.Environmental monitoring results were found to be acceptable, and they confirmed that there was not an adverse impact on environmental conditions, device quality or performance.The sterilization run (b)(4) is not related to any additional complaint infection record reported as of today.Considering that there is no adverse trend for this type of event, no additional actions are deemed required at this time.The issue with infection will continue to be monitored and corrective actions will be taken in the future if deemed appropriate.Given the fact that the cause of the infection cannot be specifically associated with the manufacturing process, and there is not an adverse trend for this type of event, and no er/ ncr(s) were identified during the investigation associated with this lot and the complaint, no corrective action is deemed necessary at this time.Dhr summary review of sterilization and seal strength records related to this work order did not identify any deviations or non-conformities that may be associated with the reported device.The sterilization process was successfully completed, and seal strength test results were found to be acceptable.(b)(4).Environmental monitoring results for august 2024, and bioburden monitoring results for august 2024 were reviewed, refer to enviro/micro summary section for more details.The review of the documentation associated with the manufacturing process indicates that all devices involved in this work order were released in accordance with abbvie¿s procedures and were no defects/problems/issues observed during the manufacturing process and the review of the documentation associated to the investigation.According to the device history record review performed, the reported event was not confirmed.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported ¿incision and drainage to obtain culture, pseudomonas aeruginosa¿.This record is for the left side.Device was explanted.
|
| |
|
Search Alerts/Recalls
|