| Model Number |
TA6048S |
| Medical Device Problem Codes |
Failure to Form Staple (2579); Firing Problem (4011); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
Tissue Breakdown (2681)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during a robotic colostomy reversal that was converted to open due to adhesions, during rectum transection before firing of the circular stapler, the staples deployed, but it did not form around the tissue.It can be seen that the staples were poking through the reload.The surgeon was unsure if the lever was pulled the second time to fire or if it pulled back and locked in place.The surgeon alluded to its closing, and then when it was fired, it felt stuck.The stapler could not be squeezed back to the handle and locked into place to sit flush with the back of the open stapler.The issue was resolved by resecting and re-stapling the tissue.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the instrument was received loaded with a 4.8mm single use loading unit.The sulu noted that all of the staples were partially advanced in the cartridge.The staples were in unusable and unformed condition.The pusher slots that did not have partially advanced staples were not loaded with any staples.The identifying witness mark from automatic firing fixture was present on the instrument handle.Functional testing found that the advanced unformed staples in the sulu were reused and reset inside the cartridge.The instrument and the sulu were applied to appropriate test media.The jaw closed smoothly, the pin slide advanced fully, and the handle actuated smoothly into pre-clamped and clamped positions.The jaw opened, handle unlocked and pin slide retracted when black release button was depressed.Acceptable staple formation was observed on test media.It was reported that the staple line was incomplete, difficult to fire, handles binding and the staple did not form.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue may occur when the firing stroke is not completed during application.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: to fire the instrument, squeeze the handle a second time until it reaches a solid stop and it locks in the back position.If the ins trument handle is partially squeezed during firing then released, bleeding may occur as the instrument was not fully fired.The black release button may be used to open the instrument at any time during approximation.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|
|
|