| Catalog Number |
51-102050 |
| Medical Device Problem Codes |
Tear, Rip or Hole in Device Packaging (2385); Device Contaminated During Manufacture or Shipping (2969)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/26/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).G2: foreign, japan.The reported product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that damaged sterile packaging with urethane identified during inspection of circulated stock items.There was no patient involvement.No additional information available for the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay additional information.The following sections were updated: g3; h2; h3; h6.Visual evaluation of the returned product/provided photos identified damage to the sterile packaging (pouch).Sterility has been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The root cause of the reported event can be attributed to transit damage and a packaging design issue.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor trends.
|
| |
|
Event or Problem Description
|
|
No additional information available for the reported event.
|
| |
|
Search Alerts/Recalls
|