SYNTHES GMBH CHUCK WITH KEY/REAM SPEED FOR TRS; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
| Catalog Number |
05.001.207 |
| Medical Device Problem Codes |
Mechanical Jam (2983); Complete Loss of Power (4015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the chuck attachment device was frozen/would not move.It was noted in the service order that the device did not work.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: a visual and functional assessment was performed on the device.The following steps failed: general condition; markings; check maximum range of tool coupling; check handpiece coupling; check pins length of handpiece coupling; check general function in running mode.During the service evaluation the following failures were identified: functional : frozen/will not move, visual : cosmetic damage - worn etch, visual : color band chipping/fading, device interaction (2+ devices) : cannot engage attachment - coupling, labeling : illegible etch.Conclusion: failure reported is confirmed root cause: component failure from wear.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: corrected data: h11.A visual and functional assessment was performed on the device.The following steps failed: general condition; markings; check maximum range of tool coupling; check handpiece coupling; check pins length of handpiece coupling; check general function in running mode.During the service evaluation the following failures were identified: functional: frozen/will not move; visual: cosmetic damage - worn etch; visual: color band chipping/fading; device interaction (2+ devices): cannot; engage attachment - coupling.Conclusion: failure reported is confirmed.Root cause: component failure from wear.
|
| |
|