| Catalog Number |
D128211 |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/01/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a patient underwent an atrial fibrillation ablation procedure with a pentaray nav catheter and during the operation, the device (including port, luer hub) was not irrigating.The catheter was primed with a pressurized bag instead of using an irrigation pump, and it was found that the heparin water in the pressurized bag could not drip out.Flushing the irrigation port with a syringe was "too stressful to flush".The device had been used on the patient prior to noting the issue.A second device was used to complete the operation.There was no adverse event reported on patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster, inc., or its employees that the report constitutes an admission that the product, biosense webster, inc.Or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Manufacturer's reference number: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
On 19-mar-2026, the bwi product analysis lab received the device for evaluation.The product investigation was subsequently completed.It was reported that a patient underwent an atrial fibrillation ablation procedure with a pentaray nav catheter and during the operation, the device (including port, luer hub) was not irrigating.The catheter was primed with a pressurized bag instead of using an irrigation pump, and it was found that the heparin water in the pressurized bag could not drip out.Flushing the irrigation port with a syringe was "too stressful to flush".The device had been used on the patient prior to noting the issue.A second device was used to complete the operation.There was no adverse event reported on patient.Device evaluation details: the device was returned to johnson & johnson medtech (j&j medtech) for evaluation.Following j&j medtech procedures, a visual inspection and irrigation test of the returned device were performed.Visual inspection revealed no damage or anomalies on the device.An irrigation test was performed, and the catheter failed the test due to the irrigation tube was found folded at the tip area.A manufacturing record evaluation was performed for the finished device number 31784241l, and no internal actions related to the reported complaint were identified.The occlusion issue reported by the customer was confirmed.The root cause of the irrigation tube folded is traced to manufacturing process.The instructions for use (ifu) contain the following information: flush the catheter with heparinized saline prior to insertion into the body.Always follow standard practices of using a continuous drip of anticoagulant fluid under pressure through the proximal luer connector when the device is in the body.This product issue will be addressed through the quality system.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
| |
|
Search Alerts/Recalls
|
|
|