Product complaint # (b)(4) this report is being submitted pursuant to the provisions of 21 cfr part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Related events captured via (b)(4).H3: investigation summary according to information received, breakage of suture was reported when the surgeon attempted to sew tissue using product code vcp346h.Three products experienced suture breakage, but the surgery was completed with another device and no patient injury was reported.The product was returned to eth for evaluation.Visual inspection determine that one open box with twenty-eight (28) unopened samples were received for analysis.Product code vcp346h.As per the sampling plan, a visual inspection was performed on thirteen samples, no defects were found on the packages.The packets were open, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no issue related to breakage sutures were observed during evaluation.A functional test was performed using an instron equipment and tensile force were above the minimum requirements.As part of ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.Although no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.Device history review a manufacturing record evaluation was performed for the finished device a manufacturing record evaluation was performed for the finished device 1094b6/vcp346hcn31 and no non-conformances / manufacturing irregularities related to the malfunction were identified.Mfg.Date: 06/16/2025 expiry date: 05/31/2030.
|