• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DJO SURGICAL; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE52

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DJO SURGICAL; SHELL, BIPOLAR, MODULAR, ENCORE, SIZE52 Back to Search Results
Catalog Number 412-02-052
Medical Device Problem Code Patient Device Interaction Problem (4001)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/19/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - due to bipolar shell came apart.
 
Additional Manufacturer Narrative
The agent reported, "bipolar shell came apart.M/86." the previous surgery and the surgery detailed in this event occurred 12 days apart.This evaluation is limited in scope as the item(s) associated with this investigation was not returned to djo surgical - austin for examination.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate that the reported device(s) was defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.A review of the device history record(s) shows that the reported component(s) used in the previous surgery met design and manufacturing requirements at the time of release for use.There were no nonconforming material reports associated with the product(s) that may have contributed to the reported event.The device(s) was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.There was ncmr# (b)(4) associated with the main part # 412-02-052, shell, bipolar, modular, encore, size52, which documents that out of a 226-part lot, (b)(4) parts were rejected due to contamination with hydroxyapatite (ha) nano.1 part coated with ha nano was mistakenly cleaned with other cocr parts, contaminating them with ha').Later, the rejected parts were recleaned and accepted after proper justification.Ha coating enhances biological fixation of implants by promoting bone growth on the implant surface.It does not affect mechanical strength, geometry, or structural integrity.Therefore, ha cross-contamination on cocr components would not cause or contribute to a bipolar shell separating customer complaint history of the reported device(s) showed no present trends or ongoing issues that need review.Inventory containment is not required as there are no indications of a product or process issue affecting implant safety or effectiveness.The revision surgery was completed successfully.The root cause of this complaint was revision surgery due to the bipolar shell coming apart.No information regarding the patient's pre-existing conditions or any activities the patient was involved in that may have contributed to the event was submitted to the complaint.There are multiple factors that may contribute to an event beyond djo surgical's control.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DJO SURGICAL
Common Device Name
SHELL, BIPOLAR, MODULAR, ENCORE, SIZE52
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758
Manufacturer Contact
stacey bonnell
9800 metric blvd
austin, TX 78758
5128346313
MDR Report Key24424090
Report Number1644408-2026-00085
Device Sequence Number6052277
Product Code KWY
UDI-Device Identifier00888912077439
UDI-Public00888912077439
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K953510
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2026
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number412-02-052
Device Lot Number000C1160
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/26/2026
Initial Report FDA Received Date02/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/16/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
402-03-282 LOT: 6113510.
Outcome Attributed to Adverse Event Required Intervention;
Patient Age86 YR
Patient SexMale
-
-