• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTUITIVE SURGICAL, INC ENDOWRIST; CADIERE FORCEPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTUITIVE SURGICAL, INC ENDOWRIST; CADIERE FORCEPS Back to Search Results
Model Number 470049-08
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/14/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The instrument involved in this complaint has been received and tested by failure analysis.The instrument was found to have a damaged grip cable.
 
Event or Problem Description
It was reported that there was a broken cable with a da vinci product.The customer reported event has no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST
Common Device Name
CADIERE FORCEPS
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer (Section G)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
Manufacturer Contact
izabel nielson
3410 central expressway
santa clara, CA 95051
4085232100
MDR Report Key24425455
Report Number2955842-2026-07327
Device Sequence Number21475282
Product Code NAY
UDI-Device Identifier00886874112250
UDI-Public(01)00886874112250(11)231130(10)K10231130(91)0001
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K150284
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 02/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number470049-08
Device Catalogue Number470049
Device Lot NumberK10231130 0343
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/06/2024
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 12/10/2024
Initial Report FDA Received Date02/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES
Patient SexUnknown
-
-