| Model Number |
LUCAS 3 |
| Medical Device Problem Code |
Electrical /Electronic Property Problem (1198)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/18/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
A customer contacted stryker to report a non-critical issue with their device.Upon inspection, it was observed that the device alarmed and paused.In this state, a delay in cardiopulmonary resuscitation may occur.If a device malfunctions, the device operating instructions indicate that the user is to remove the device and immediately start manual chest compressions.There was no patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
A stryker service technician evaluated the customer's device and observed that the device alarmed and paused.After replacing the compression module, proper device operation was observed through functional and performance testing.The device was subsequently returned to the customer for use.
|
| |
|