| Model Number |
453964 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/20/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion in a severely tortuous segment of the ostial cx.While inserting the orsiro mission, the stent buckled on itself, still on the balloon.After removal from body, the stent is still intact on balloon.A stent from another manufacturer was used to complete the intervention.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes.
|
| |
|
Additional Manufacturer Narrative
|
|
Combination product: yes the returned product was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed that the stent is deformed at its proximal end.Several stent struts are deformed (i.E., crushed, compressed, everted).Microscopic analysis revealed stent imprints on the exposed balloon surface, indicating that the bent struts were initially crimped on the balloon.Outside of the deformed zone the crimped diameter of the stent complies with the specification.Review of the production documentation confirmed that the product was manufactured according to specifications and passed all in-process and final inspections.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause could be determined.
|
| |
|
Event or Problem Description
|
|
The orsiro mission drug-eluting stent system was selected for treatment of a severely calcified lesion in a severely tortuous segment of the ostial cx.While inserting the orsiro mission, the stent buckled on itself, still on the balloon.After removal from body, the stent is still intact on balloon.A stent from another manufacturer was used to complete the intervention.
|
| |
|