| Model Number |
Mi1250 SYNCHRONY 2 PIN |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/17/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
User has no speech understanding with the implanted device on the left side (bilaterally implanted).
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been explanted.If it should be explanted, it should be returned to the manufacturer for evaluation.When available, a device failure analysis will be submitted as a follow up report.
|
| |
|
Event or Problem Description
|
|
User has no speech understanding with the implanted device on the left side.Reimplantation will be scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: according to the information received from the field a damage to the reference electrode seems likely.However, to determine an exact root cause device investigation would be necessary.Re-implantation is considered but no date has been scheduled yet.
|
| |
|
Event or Problem Description
|
|
User has no speech understanding with the implanted device on the left side.The user was reimplanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Conclusion: damage to the reference electrode, likely caused by an external impact, was determined to be the root cause of device failure.Other mechanical damages found are attributable to the removal surgery.This is a final report.
|
| |
|