Correction - d9, h6 method code.The reported event could not be confirmed based on the evaluation of the returned devices and the available information.The device inspection revealed the following:
visual examination of the returned device shows scratches and blemishes consistent with the explantation process.The tibial tray was received assembled with the poly and could not be separated upon inspection.Residual bone cement and areas of ossification are present on both the tibial tray and talar dome.Overall, the device appears to be in good condition.Additionally, key measurements were taken with a digital caliper and found to be within dimensional specifications.Failure of total ankle replacement (tar) over time is a known complication.In the case of a planned revision procedure, a medical opinion aimed at determining the root cause of the failure is part of the internal investigation process.Sufficient radiological and clinical information must be provided to enable a meaningful clinical assessment and to identify possible causes of failure.In cases where a ct scan is the only available clinical source, this assessment is limited.No conclusive statement can be provided because key clinical information (e.G., clinical status, exact symptoms and range of motion of the patient, and assessment of the treating physician) is missing.The root causes of radiographic findings such as radiolucent areas and bone cysts are often complex and multifactorial and cannot be determined scientifically without this critical evidence.Due to these limitations, we are unable to provide statements regarding the patient, the procedure, or the device in relation to the cause of the failure.More detailed information about the complaint event must be available in order to determine the root cause of the complaint.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing, or design-related problems were found during the investigation.If additional information is provided, the case will be reassessed.
|