|
Product complaint # (b)(4).Date sent to the fda: 2/24/2026.H6.Component code: c22 - photo analysis.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Additional information was requested, the following was obtained: when were the needles detached/pulled off from the suture (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Received information as following from sales rep via phone today.Please refer to the event description, other information requested is unknown.H3.Investigational summary: the product was returned to ethicon for evaluation.One open box with twenty-one unopened samples was received for analysis.The product code is vcp213h.Two photographs were provided.The first depicts two boxes labeled with product code vcp213h.The second depicts one box and some winding formers in two plastic bags.Upon visual inspection of the returned samples, the swage and attachment area were noted to be as expected.The sutures were dispensed without any problem.The sutures were noted with significant difference in suture diameter and coloration at the tipping area.As per this condition observed in the sutures were submitted to flex test.The samples broke due to improper tipping, as the sutures were breaking away from the swage area.Based on the information currently available, damaged suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.Additional h3 investigational summary: the product was returned to eth for evaluation.Visual inspection revealed that one winding former with a needle and one dispensed suture outside the foil packet were received for analysis.Product code vcp213.The swage and attachment area were noted as expected during the visual inspection of the needle.The barrel hole of the needle was examined under magnification and remnant suture was found.The suture was examined, and although the diameter of the suture in the tipping area was noted with significant difference in suture¿s diameter and coloration.Besides that, the end was observed to be broken.This type of detachment mode is most likely related to improper tipping, as the suture is breaking away from the swage area.Based on the information currently available, damaged suture was identified during the investigation of the sample received.This product issue will be addressed through the ethicon quality system.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of post-market surveillance.
|