| Catalog Number |
305122 |
| Medical Device Problem Code |
Contamination (1120)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/10/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.B3.The date received by manufacturer has been used for this field.
|
| |
|
Event or Problem Description
|
|
It is reported foreign matter.Description: not able to use bent has a couple of one that has contamination, metal shavings.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4) follow up.It was reported the needle was bent and a couple have metal shavings.Our quality team has completed a device history record review for material number 305122 and lot number 3320178.The review did not identify any abnormalities during the production process that could have contributed to the condition, and all quality tests were within specification.Due to the unavailability of a sample for return, a comprehensive sample investigation could not be conducted.Consequently, the exact cause of this incident remains undetermined.Should you encounter any further issues with our product, we would appreciate the opportunity to perform a detailed analysis.At this time, no further action is deemed necessary.
|
| |
|