While the customer was using a part of an intraoral sensor system, schick33 s2 sensor ship kit/3.0 cable 9', they allege the device self triggered, causing an x-ray image to be taken without prompting the device to take an image.No injury was reported from the alleged event.
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation: problem description: (b)(6) 2026, 12:06:41, (b)(4).Initial log: (b)(6) 2026 11:57:54 am (dentsply (b)(6) time).Warranty: warranty.Dealer on-site: no.Billable: not billable-warranty.Billing accepted: na.Billing approved by: na.90 day expiration date: 05/03/2026.Remote access: (b)(4).Sidexis version: none.Problem description: tried to replace the sneosr cables and it wont work.Tried different usb module on two different pcs.All other sensors work fine.Troubleshooting description: tv connection.The sensor auto fires when trying to ready the device.Solution description.(b)(6) 2026, 16:57:30, (b)(4).Initial log: (b)(6) 2026 12:05:37 pm (dentsply (b)(6) time).Solution description: recommended to replace the sensor.Recommended parts: 100008652.Office has not used this sensor to take images on a patient yet.No patient was injured.
|