While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- warranty: yes.Dealer on-site: no.Billable: billing accepted: billing approved by: 90 day expiration date: 05/04/2026.Interface connection issue: solution description: 02/03/2026, 09:22:20, (b)(4).Interface connection issue: remote access: no, sidexis version: server location: pdata remote location: rcu location (if applicable): practice management: workstation(s): scout check: firmware: plug-in version: exposure count: unit ip address: troubleshooting details: interface not working.The office has tried multiple interfaces with different sensors, and they are working.Sending a replacement interface functional test successful / follow up: 100008290 schick 3.0 usb interface, rma.Note: 02/09/2026, 10:59:09, (b)(4).Additional information: checked with (b)(4), no injuries reported, closing ticket.
|