ABBOTT DIABETES CARE INC FREESTYLE LITE; BLOOD GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 70805-70
Medical Device Problem Code Failure to Power Up (1476)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 02/12/2026
Type of Reportable Event Malfunction
Event or Problem Description
The reporter contacted abbott diabetes care, alleging the meter will not turn on.This complaint is being reported because the reported issue was not resolved with troubleshooting.There was no report of death, serious injury, or mistreatment associated with this event.
 
Additional Manufacturer Narrative
The product has been requested back for an investigation and the customer agreed to return the device; however, at this time product has not yet been received.The most probable root causes associated with this failure mode are disconnected, faulty or damaged components, software/data corruption, or misuse.However, mitigations are in place to reduce and prevent such issues.All vital manufacturing steps are validated, monitored, and verified during manufacturing to ensure the system is in conformance with the verified design.All vital functions are monitored by the system and, when necessary, function is suspended to safeguard against inaccurate results.Labeling is provided to instruct the user on the intended use of all vital parts of the system to minimize misuse.All complaints and complaint trends are investigated to determine if there is a product defect/ deficiency.If a product defect/ deficiency is identified, a risk evaluation is completed and compared to the risk management report, to ensure the risk profile has not changed.Additionally, as a part of abbott¿s post-market surveillance process, all risk evaluations with associated complaint data are reviewed annually to determine if the risk profiles have changed as compared to the product risk management reports.These monitoring processes ensure that all product risk profiles remain acceptable and have a positive benefit/ risk ratio.An extended investigation has been conducted, which involved a manufacturing review (including 5 years of device history records (dhrs)), previous complaint and corrective and preventive action (capa) investigations conducted for blank screen issues, process failure mode effects analyses (pfmeas), design controls, and design failure mode effects analyses (dfmeas), risk management reports, risk evaluations, and label copy.The investigation did not identify any indication that the product did not meet specification.If product is returned, a physical investigation will be performed per adc's established processes and procedures and a report will be submitted upon completion of investigation.All pertinent information available to abbott diabetes care has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FREESTYLE LITE
Common Device Name
BLOOD GLUCOSE MONITORING SYSTEM
Manufacturer (Section D)
ABBOTT DIABETES CARE INC
1360 south loop road
alameda 94502 7001
Manufacturer Contact
patricia reis castro
1360 south loop road
alameda 94502-7001
4157205405
MDR Report Key24429655
Report Number2954323-2026-10501
Device Sequence Number10227515
Product Code NBW
UDI-Device Identifier00699073708045
UDI-Public(01)00699073708045(10)02C082D(17)260831
Combination Product (Y/N)N
PMA/510(K) Number
K092602
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 02/24/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number70805-70
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/12/2026
Initial Report FDA Received Date02/24/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/26/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexFemale
-
-