| Model Number |
EEAORVIL25A |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/23/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that during a laparoscopic roux-en-y gastric bypass procedure, while on anastomosis of stomach to the small intestine, the trocar and anvil failed to attach and the metal tip was open and not secure when inserted into the stomach.The device was pulled out and a new orvil was inserted to resolve the issue.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the orvil anvil was observed to be tilted.Further inspection of the orvil anvil cutting ring showed no traces of knife impression and the ring was received intact.One of the retainer legs of the orvil anvil was observed to be bent.It was reported that the trocar and anvil failed to attach and the metal tip was open and not secure when inserted into the stomach.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue can occur if the orvil anvil is manipulated with excessive force during the attachment to the instrument, or if the orvil anvil is mishandled during the removal from the tubing.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: do not grasp/clamp on the legs (open end) of the anvil/center rod assembly.Doing so could bend the legs and make it difficult to at tach/detach the anvil to the integrated trocar of the stapler.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|