| Catalog Number |
6778240 |
| Medical Device Problem Code |
Poor Quality Image (1408)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description (b)(6)2026 09:49:58 rafael kowalski (b)(6) sensor will be replaced note (b)(6) 2026 02:09:04 rafael kowalski (divxu690) according to technicians, the patient sustained no injuries.
|
| |
|
Event or Problem Description
|
|
While the customer was using a part of an intraoral sensor system, xios xg supreme ae usb s2 sensor,270cm, they allege the device self triggered, causing an x-ray image to be taken without prompting the device to take an image.No injury was reported from the alleged event.
|
| |
|