| Model Number |
V5 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Abdominal Pain (1685); Bowel Burn (1756); Fever (1858); Uterine Perforation (2121)
|
| Date of Event |
02/23/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported during a novasure procedure on (b)(6) 2026, that the device was initially seated and the cavity integrity test passed.The ablation began and the ablation time was 1 minute and 8 seconds when the vacuum alert displayed and the ablation was stopped.It was then verified that the tubing was hanging below the patient's leg.No leak was noted at the cervix, no fluid was overflowing the tubing to the desiccant, and the vacuum valve was actuated.The vacuum alert continued and that's when the physician chose to try another device.After seating the array on the second device, the cavity assessment test failed and a hysteroscopy was performed.A perforation at the fundus was noted and the physician then performed a laparoscopy to repair and suture the perforation, and the surgery was successful.The patient is waking up from surgery.No additional injury noted to the surrounding structures.No further surgery or medication was administered to the patient.No other information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Lot number of the device not provided by the complainant; therefore, the udi, expiration and manufacturing dates are not known.Device history record (dhr) review was unable to be conducted for the disposable device as the identification numbers were not provided by the complainant.The device involved in this event was not returned for evaluation purposes therefore visual and functional analysis of the product could not be performed.We are unable to confirm a relationship between the device and the issue reported.The information obtained during complaint investigation will be included in our global complaint trending and product surveillance will continue to monitor complaints of this type for adverse trends.If the product is received or additional information is obtained, the investigation will be reopened accordingly per standard operating procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received: the patient was experiencing abdominal pain and had a fever so the patient went to emergency (er).Upon examining the patient, it was found that the patient had a bowel burn.A bowel resection was then performed.The patient is still in the hospital and recovering.
|
| |
|
Search Alerts/Recalls
|