| Model Number |
H938174 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
E1: initial reporter postal code: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that particulate matter (pm) contamination was observed on the tubing and in the bag of three (3) vented, high-volume inlets.The pm was described as ¿black spots on the tubes and in the bag¿.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, h3, h4, h6(update codes), h11.H4: the lot was manufactured in december 2024.H11: two (2) devices were received for evaluation.Visual inspection on the returned samples identified a dark spot embedded outside the tubing and not in the fluid pathway, as well as a white spot embedded inside the transparent cap of the white spike connector and within the fluid pathway.The reported condition was verified.The cause of the issue could not be determined; however, the cause was possibly inherent to contamination during the manufacturing or packaging process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|