While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Solution description: 02/06/2026, 14:14:05, (b)(4).Solution description: tech swapped elastomer, sensor cable and remote with usb cable and found the issue is the sensor.Replace sensor.Functional (yes/no): troubleshooting description: they found the artifact during taken an exposure on a patinet.They followed up with blank test exposure and found the issue was the sensor no patient was injured.Just a bad sensor.
|