While the customer was using a part of an intraoral sensor system, schick ae usb interface, they allege there was an x-ray image transmission issue where the x-ray image was not captured after radiation has occurred (transmission failure).No injury was reported from the alleged event.
|
|
While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- problem description (b)(6) 2026, 14:21:16, (b)(6).Initial log- (b)(6) 2026 4:16:09 pm (b)(6).Warranty: yes dealer technician on-site: no billable: no (equipment under warranty) remote access: sidexis version: server location: pdata remote location: rcu location (if applicable): practice management: workstation(s): scout check: firmware: plug-in version: exposure count: unit ip address: troubleshooting details: the usb interface is not capturing images when the sensor is exposed to radiation when they try to retake images, some are able to process and others are not sent to sidexis recommended to have the usb interface replaced this occurs with and without patients no injuries reported.Replaced part.
|