| Catalog Number |
ES2820 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
Pain (1994)
|
| Date of Event |
01/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the implant was removed due to pain and an unspecified malfunction.
|
| |
|
Event or Problem Description
|
|
According to the available information, the implant was removed due to pain and an unspecified malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Titan touch pump, cylinders 1 and 2, and reservoir were received for evaluation.Abrasion was noted on both exhaust tubes and the inlet tube of the pump.No functional abnormalities were noted with the pump.Abrasion was noted on the exhaust tubes of cylinder 1 and cylinder 2.A separation was noted on the exhaust tubing of cylinder 1, near the strain relief junction.This is a site of leakage.The surfaces appear to have crease marks adjacent to or within a confined area of abrasion, indicating the area was kinked and abrading against itself for an extended period of time.No functional abnormalities were noted with cylinder 2.Abrasion was noted on the inlet tube of the reservoir.A separation was noted on the inlet tube of the reservoir near the strain relief.This is a site of leakage.A small area of abrasion was noted adjacent to the separation indicating the tube had been kinked onto itself and abraded.Based on examination of the returned product, it was concluded that the abrasion marks noted on all the tubing matched in such a way to conclude that they had overlapped and abraded against one another while in-vivo.This then may have caused the exhaust tube of cylinder 1 and the inlet tubing of the reservoir to be kinked onto itself for a period of time to cause the observe separations.This positioning, in combination with device usage over time, could contribute to sufficient stress to separate the exhaust tubing and inlet tubing.Separations of this type could then allow the loss of fluid, making the device inoperable.A review of the device history record confirmed the devices from this lot met all specifications prior to release.A review of the complaint history database, nonconformances and capas revealed no trends for this lot.
|
| |
|
Search Alerts/Recalls
|