While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- problem description: 02/17/2026, 10:47:43, (b)(4).Usb 3.0 connection issue.Solution description: 02/17/2026, 10:47:43, (b)(4).Did the reported issue require multiple images / multiple exposures to be taken on a patient: yes.Were there any injuries or misdiagnosis if multiple images/exposures were needed: no.Ds ticket # or troubleshooting: issue reported: remote not working intermittent did the reported issue require multiple images / multiple exposures to be taken on a patient (yes/no - do not put na) y.Were there any injuries or misdiagnosis if multiple images/exposures were needed: no.Equipment type and model: ae remote.Serial: number: (b)(6).Confirm sensor cable color: gray acquisition software & version: es 23.Issue in one or all rooms: all.Other sensors work with working remotes in room(s) with issue: y.Pc name: shawddm0020.Schick driver version (programs and features): remote fw/fpga version: na.Sensor fw version: if remote issue: 2.0 or 3.0: tried different usb cable/usb port/bypassed usb hubs/extenders: y.3.0 remote: tried on a 3.0 and 2.0 usb port/cable clip connected: n.If sensor issue, replaced elastomer: if sensor issue, tried different sensor cable: if no response to radiation, confirmed xray head functional: additional troubleshooting steps/notes: replaced part.
|