|
|
| Catalog Number |
HAR723 |
| Medical Device Problem Codes |
Material Separation (1562); Failure to Read Input Signal (1581)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/11/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/24/2026.D4 batch #: unknown.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that during a hepatectomy the device successfully passed the preoperative test.At some point the surgeon activated the device and the generator displayed the message release pressure on blades.He then noticed that the active blade was broken.The surgeon replaced the device with another har723 and removed the broken blade, after which the procedure proceeded without further complications.There were no patient consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).Date sent: 2/27/2026.D4 batch #: y70184.Investigation summary- the product was returned for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that the device was returned with the distal tip of the blade broken off and not returned.The remaining blade portion was scratched, and with evidence of contact with metal in or out of the operative field.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was disassembled to verify the internal components and no anomalies were found.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure or using any means other than the blade wrench to attach or detach the blade.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result are such as ¿tighten assembly¿, ¿blade error detected¿ or "relaxed pressure on blade" followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.As part of the quality process all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch y70184, and no non-conformances were identified.
|
| |
|
Search Alerts/Recalls
|
|
|