While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- note (b)(6) 2026, 11:23:30, (b)(6).Further action further action (b)(6) 2026 ,11:17:21 am (b)(6).Caller's full name: (b)(6) caller's phone number: caller's role: office staff.Warranty status: billing previously accepted: problem description: cable was replaced and sensor still having issues are injuries reported? if yes, what type? no was xray radiation emitted during reported issue? yes, was a patient exposed to radiation? if yes, how many times?2 stopped in the middle of a bitewing series.Are example images attached:no troubleshooting description: recommend replacing sensor.
|