• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; PROSTHESIS, PENIS, INFLATABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AMS 700 CXR; PROSTHESIS, PENIS, INFLATABLE Back to Search Results
Model Number 72404013
Medical Device Problem Codes Degraded (1153); Mechanical Problem (1384); Nonstandard Device (1420)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/16/2026
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This product was manufactured prior to implementation of the udi regulation and as a result, the complete udi is not available.Applicable us product data has been included in this report.The device is not available for analysis; therefore, no physical analysis of the product could be performed.The reported device performance allegation cannot be confirmed.Mechanical issue and bulge are acknowledged within the dfu and are not related to off-label use or failure to follow instructions.Additionally, it is concluded that the cause of the allegation cannot be established without product return.Based on the information available the cause that contributed to the reported event cannot be established.
 
Event or Problem Description
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with a nonfunctioning device.A revision surgery was performed, during which the physician confirmed that the right cylinder had a ballooning.The device was explanted and replaced.There were no patient complications.
 
Additional Manufacturer Narrative
Model number/catalog number: 72400152.Serial number: n/a.Batch/lot number: 671122002.Model/catalog description: reservoir.Model number/catalog number: 72401110.Serial number: n/a.Batch/lot number: 644410003.Model/catalog description: pump cxm.This product was manufactured prior to implementation of the udi regulation and as a result, the complete udi is not available.Applicable us product data has been included in this report.Upon receipt at the quality assurance laboratory, the returned components underwent evaluation, and the first cylinder exhibited a bulge during decontamination, which was punctured with a sharp instrument to release contaminated fluid.When inflated with a syringe, the first cylinder exhibited a bulge consistent with excess air between the inner and outer tubes and exhibited wear at a fold in the proximal end of the cylinder, with a hole identified in the outer tube consistent with kink resistant tubing (krt) wear.Additional wear was observed in the middle and distal sections, along with broken and detached fabric threads in the proximal end, and a hole and leak were identified in the middle of the cylinder, attributed to sharp instrument damage.The second cylinder exhibited wear at a fold in the proximal end of the cylinder, with additional wear observed in the middle and distal sections, and the second cylinder passed leakage testing.The standard inflate deflate pump exhibited krt worn to filament, and the outer layer of the pump bulb exhibited wear against the pump strain relief krt junction, with the krt exhibiting a hole and leak consistent with fatigue.The spherical reservoir exhibited wear at folds in the reservoir shell, with multiple folds observed, and the reservoir passed leakage testing.The device history record (dhr) review confirmed that the device met all material, assembly, and performance specifications.A review of the device instructions for use (ifu) was completed and did not reveal any evidence of device misuse, off label use, or failure to follow instructions.A risk review was completed and confirmed that the events of mechanical malfunction (leaks, incomplete inflation/ deflation, kinking), device has a fluid leak, and inconsistent cylinder diameter (aneurysm) were defined in the product risk documentation.This event type has been accounted for during analysis to support acceptable risk benefit for the product.Based on the information available and the analysis results, the first cylinder was confirmed to have a bulge during decontamination, and the bulge was punctured with a sharp instrument to release the contaminated fluid as it did not drain on its own.The cylinder bulge would lead to a mechanical issue as the cylinder could no longer inflate or deflate correctly.A conclusion code of cause traced to component failure was assigned to this investigation.
 
Event or Problem Description
It was reported that a patient with an inflatable penile prosthesis (ipp) presented with a nonfunctioning device.A revision surgery was performed, during which the physician confirmed that that the right cylinder had a ballooning.The device was explanted and replaced.There were no patient complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMS 700 CXR
Common Device Name
PROSTHESIS, PENIS, INFLATABLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer Contact
farshad fahimi
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key24431755
Report Number2124215-2026-10170
Device Sequence Number10297566
Product Code JCW
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
N970012
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2026
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/19/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number72404013
Device Catalogue Number72404013
Device Lot Number1100575619
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2026
Supplement Date Received by Manufacturer04/21/2026
Initial Report FDA Received Date02/24/2026
Supplement Report FDA Received Date05/19/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/22/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PUMP LOT 644410003, UPN 72401110; PUMP LOT 644410003, UPN 72401110.; RESERVOIR LOT 671122002, UPN 72400152; RESERVOIR LOT 671122002, UPN 72400152.
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age84 YR
Patient SexMale
-
-