While no serious injury resulted in this event, if this malfunction recurred, it could cause or contribute to a serious injury or require medical or surgical intervention to preclude such.This event, therefore, is reportable per 21cfr part 803.Usb 3 firmware v1.4 contains an issue in the portion of code used to optimize sensor performance.If the optimization occurs during an exposure or sensor readout, there is a chance that the resulting image will be corrupted.Image data following the point of corruption will consist of data that was previously held in the memory of the usb device.As a result, the presented image may contain a portion of data from a previous exposure.In general, this would be obvious to the user since the presented image would either contain obvious artifacts or different anatomy than expected.Field service evaluation- solution description 02/18/2026 14:12:49 (b)(6).Initial log 02/18/2026 11:31:59 am (dentsply sirona charlotte time).Warranty: 04/22/2024 dealer on-site: billable: n canada billing accepted: billing approved by: 90 day expiration date: 05/19/2026.Remote access: sidexis version: 4.31 server location: \\tfd-server\pdata pdata remote location: rcu location (if applicable): practice management: tracker workstation(s): tfd-op2-924, tfd-op3, tfd-op4, tfd-op1 scout check: firmware: version 1.5.140 version 3.4 plug-in version: ae interface upgraded to ioss 3.2 exposure count: unit ip address: troubleshooting description: i remoted into tfd-op2-924 and removed ae interface driver and replaced it with ioss 3.2 scrambled images.I remoted into tfd-op3 i uninstalled ae interface driver and installed ioss 3.2 i remoted into tfd-op4 i uninstalled ae interface driver and installed ioss 3.2 i remoted into tfd-op1 i uninstalled ae interface driver and installed ioss 3.2 scrambled images can be due to elastomer.I provided part number for cable replacement to replace cable and elastomer.(b)(4).Sensor with scrambled images.Patient involved 4 times the image has been retaken no injuries.
|
|
While the customer was using a part of an intraoral sensor system, schick33 s2 starter kit/3.0 interface 9', they allege experiencing image quality issues when an x-ray image was taken and an additional image was required.No injury was reported from the alleged event.
|