• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; COLONOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CRE WIREGUIDED; COLONOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID Back to Search Results
Model Number M00558400
Medical Device Problem Codes Burst Container or Vessel (1074); Leak/Splash (1354)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/28/2026
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a0402 captures the reportable event of balloon burst.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was being prepared to be used in the common bile duct in a duodenoscopic choledocholithotomy procedure performed on (b)(6) 2026.It was indicated that the procedure was performed because the patient had bile duct stones with cholangitis, abdominal pain, fever and obstructive jaundice.During preparation, when the balloon was filled with water, it was found that the balloon had ruptured and was leaking.No piece of the balloon detached inside the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Event or Problem Description
It was reported to boston scientific corporation that a cre wireguided dilatation balloon was being prepared to be used in the common bile duct in a duodenoscopic choledocholithotomy procedure performed on (b)(6) 2026.It was indicated that the procedure was performed because the patient had bile duct stones with cholangitis, abdominal pain, fever and obstructive jaundice.During preparation, when the balloon was filled with water, it was found that the balloon had ruptured and was leaking.No piece of the balloon detached inside the patient.The procedure was completed with another of the same device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be stable.
 
Additional Manufacturer Narrative
Block h6: imdrf device code a0402 captures the reportable event of balloon burst block h11: investigation results the returned cre wireguided dilatation balloon device was analyzed, and a visual inspection found no damages.A functional inspection was performed, and the balloon was inflated without any problem; however, it was not able to hold the pressure due to a pinhole in the balloon which produces a leakage, located approximately at 25 mm from the tip of the device.Microscope inspection confirmed the presence of a pinhole.No other problems with the device were noted.With all available information, boston scientific concludes the reported event of balloon burst was not confirmed.The pinhole problem found could have been interpreted by the customer as the reported event of balloon burst.The results of the analysis performed on the returned device found that the balloon had a pinhole at located approximately at 25 mm from the tip of the device, which causes the reported leak.It is possible that the pinhole found in the balloon occurred due to procedural factors such as excess of pressure, interaction with other devices, or anatomical conditions.Also, it is possible that interaction with a sharp surface during/previous the procedure, could have caused the failure found on the distal section of the balloon.Based on a thorough review of the reported complaint, boston scientific has assigned an investigation conclusion code of adverse event related to procedure a review of the device history record (dhr) was performed.The review of the dhr identified that there were no process related non-conformances, scrap, or rework performed during the production that could explain the event.The reviews ensure each device meets specification prior to release for use.There is no indication the device manufacturing process contributed to the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE WIREGUIDED
Common Device Name
COLONOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork T12 Y K88
EI   T12 YK88
Manufacturer Contact
christina lam
300 boston scientific way
marlborough, MA 01752
9145716918
MDR Report Key24434589
Report Number3005099803-2026-00718
Device Sequence Number21406893
Product Code FDF
UDI-Device Identifier08714729339335
UDI-Public(01)08714729339335(17)260703(10)32149568
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K110833
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/03/2026
Device Model NumberM00558400
Device Catalogue NumberM00558400
Device Lot Number0032149568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/20/2026
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/29/2026
Supplement Date Received by Manufacturer03/12/2026
Initial Report FDA Received Date02/24/2026
Supplement Report FDA Received Date04/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/03/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age73 YR
Patient SexFemale
Patient Weight60 KG
-
-