| Model Number |
TCDS0801-XT |
| Medical Device Problem Code |
Difficult to Flush (1251)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/05/2026
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device remains implanted in patient.The device will not be returned for evaluation.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2026, a patient presented with grade 5 functional tricuspid regurgitation (tr), enlarged atrium and large leaflet gap for a triclip procedure.During the procedure, first triclip was implanted on medial side of anterior and septal leaflet (a/s).Second triclip was implanted on the central side if anterior and septal leaflet (a/s).Third triclip was implanted on the central side of anterior and septal leaflet (a/s).After releasing the third clip, it was determined that a single leaflet device attachment (slda) has occurred and clip was no longer attached to the anterior leaflet.Fourth clip was implanted on the posterior leaflet 2 and septal leaflet (p2/s).Fifth clip was implanted on posterior leaflet 1 and septal leaflet (p1/s) to stabilize the slda.It was reported that during preparations of all of the cds devices it required additional troubleshooting of advancing and retracting the device multiple times at a consistent and slow place to thoroughly de-air the system.The final tr was grade 1-2.There were no adverse patient effects or clinically significant delays reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot-specific quality issue.Based on the available information, the cause of the reported difficult to flush was unable to be determined.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|