MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG BIOMONITOR III; IMPLANTABLE CARDIAC MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG BIOMONITOR III; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number 436066
Medical Device Problem Code Failure to Sense (1559)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/21/2026
Type of Reportable Event Serious Injury
Event or Problem Description
Loop was explanted due to no sensing.No adverse patient events were reported.Should additional information be received, this file will be updated.
 
Additional Manufacturer Narrative
The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated and all production steps were performed accordingly.In particular, the final acceptance test proved the device functions to be as specified.The available data were thoroughly inspected.The recorded secgs documented a loss of signal since august 09, 2024.The root cause of these observations was not determinable.However, it cannot be excluded that external influences such as strong electromagnetic fields, like an mri, contributed to the observation.During the further course of the analysis the device could be re-initialized successfully without any difficulties.Afterwards the device was working as specified.At a next step the device was subjected to an electrical analysis.A sensing test was performed and the device sensed the attached heart signals free of noise, proving the sensing functions to be as specified.The analysis revealed no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOMONITOR III
Common Device Name
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
Manufacturer Contact
6024 jean road
lake oswego, OR 97035
8772459800
MDR Report Key24435247
Report Number1028232-2026-00964
Device Sequence Number13338085
Product Code MXD
UDI-Device Identifier04035479158453
UDI-Public(01)04035479158453(17)231231
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K190548
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2026
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/09/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2023
Device Model Number436066
Device Catalogue Number436066
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2026
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/13/2026
Supplement Date Received by Manufacturer04/08/2026
Initial Report FDA Received Date02/24/2026
Supplement Report FDA Received Date04/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/15/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient Age73 YR
Patient SexMale
-
-