| Model Number |
VTICM5 13.2 (-13.00/2.5) |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation: 4110 - lens work order search: no similar complaint type events reported for units within the same lot.Manufacturing narrative: secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
A facility representative reported that an implantable collamer lens was implanted into the patient's eye.The patient experienced refractive surprise.The lens was exchanged with a same length but different power lens and the problem was resolved.The cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 - device evaluation: lens was returned in a microcentrifuge vial, in liquid and residue/debris on product.Visual inspection found the lens haptic torn.Dimensional and functional inspection found the lens to be within specifications.Claim # (b)(4).
|
| |
|