| Model Number |
E160 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
02/04/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed that this device battery voltage has recently began dropping in an irregular fashion.Device replacement was recommended.To date, this device remains in service.No adverse patient effects were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that this implantable cardioverter defibrillator (icd) recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.A request was made to have data from this device analyzed.Data analysis confirmed that this device battery voltage has recently began dropping in an irregular fashion.Device replacement was recommended.To date, this device remains in service.No adverse patient effects were reported.Additional information indicated that this icd was explanted and replaced with a new device of different model.No additional adverse patient effects were reported.
|
| |
|