|
OLYMPUS WINTER & IBE GMBH TELESCOPE, 3 MM, 12°, AUTOCLAVABLE; RIGID ENDOSCOPE TELESCOPE, HYSTEROSCOPES
|
Back to Search Results |
|
| Model Number |
A4673A |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.A root cause could not be identified.Based on the results of the investigation, the most probable cause is traced to component failure.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
It was observed that during the device evaluation, the rigid endoscope telescope exhibited a loose meniscus.There was no patient involvement.
|
| |
|
Search Alerts/Recalls
|
|
|