| Model Number |
CD1411-36Q |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Unspecified Infection (1930)
|
| Date of Event |
02/19/2026
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the patient presented with an elevated white blood cell count and was found to have vegetation on the right ventricular (rv) lead via transesophageal echocardiogram (tee).The implantable cardiac defibrillator (icd) and the rv lead were explanted due to the infection.The patient was in stable condition.
|
| |
|
Additional Manufacturer Narrative
|
|
A device history record (dhr) review was performed and all required manufacturing processes and inspection steps were confirmed to be completed per the requirements.The device met specifications prior to leaving abbott manufacturing facilities.Review of the sterilization records confirmed normal sterilization cycles for the products.The product was returned and visual inspection was normal.Interrogation of the device revealed it was above elective replacement indicator (eri) when received.The cause of infection could not be traced to the device.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: section a1.The correct patient identifier should have been (b)(6), rather than (b)(6).
|
| |
|