|
Tw (b)(4).Updated sections: b4, d9, g3, g6, h2, h3, h6, h11.The device was returned to the factory for evaluation on 03/02/2026.An investigation was conducted on 03/03/2026.A visual inspection was conducted.Signs of clinical use and no evidence of blood was observed.There were no visual defects observed on the intact endoscope.An image quality inspection was performed with an endoscopic video imaging system.A clear image was able to be obtained.Based on the returned condition of the device as well as the evaluation results, the reported failure "poor quality image" was not confirmed.Reported serial # (b)(6).A lot history search is not applicable because this is a reusable, oem device.Due to the age of the device, a c of c is not readily available on-site.
|